Identifying Errors and Troubleshooting
en
17.13 Unexpected Changes in Functionality
Sudden, unexpected changes in the functionality or performance of components could
be caused by strong electromagnetic radiation.
Electromagnetic radiation as the cause
•
Are other electronic devices in the patient environment active, for example as part
of therapy?
•
Is the patient in a different environment?
•
Do the changes in functionality occur during mobile use?
Corrective actions
•
Remove the sources of interference from the patient's environment. Observe the
minimum distances to other electronic devices, see chapter 23: EMC, page 201.
•
If the source of interference cannot be removed in the short term, ensure that the
patient is being supplied adequately. Monitor the correct filling and emptying of
the blood pump. Monitor patient vital parameters.
17.14 Driving Tube Repair
A defective driving tube must be replaced immediately. If a new driving tube is not im-
mediately available, the defective driving tube can be temporarily repaired. The goal is
•
to seal leakages in order to maintain the flow
•
to prevent the penetration of foreign bodies into the blood pump and driving unit
Missing or illegible inscriptions or safety
symbols
Notify contact person.
Display failure
Replace the driving unit.
Display frozen, activity symbol showing
Failure of backlight
<Button>
on driving unit does not
respond or is permanently on
Changes to audible alarms indicate fail-
ure of an alarm module. Failure can be
identified through changes in alarm vol-
ume or excessively long pauses
between alarms. Audible alarms must
be emitted as described in
section 17.3.7: Audible Alarms on the
Driving Unit, page 158.
<Charge level button>
on battery does
not respond or is permanently on
Replace the battery
Error
Removal
Tab. 17-7
Errors without messages
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...