114
1015002x00
Configuring and Operating the Driving Unit
en
WARNING
8.9
Replacing and Upgrading the Blood Pump
Heed the instructions for use on EXCOR sterile components!
Before switching to a different blood pump size, check whether a differ-
ent flow sensor will be required. Information on which sensors can be
used with which pumps and cannulae is available here: Tab. 5-1 ,
page 44. Contact the hotline regarding any questions.
Each time after replacing a blood pump, check that the new blood pump
is filling and emptying completely. Adjust parameters if necessary.
Replacement of the blood pump is only described here in terms of the work steps to
be performed on the driving unit and panel PC.
8.9.1
Replacing Blood Pumps - BVAD Left, LVAD, RVAD
1.
Initial state:
The system is in
A
UTO
operation. The panel PC is connected to
the driving unit. A sterile environment has been prepared. The blood pump
has been prepared. The necessary materials are ready for use.
2.
Log in with the
E
XPERT
user profile.
3.
Remove the cable wrap.
4.
Remove the flow sensor from the cannula. If alarms are emitted with regard to
flow or a flow sensor not being connected, press
A
UDIO
PAUSED
. After one
blood pump has been paused, audible alarms will no longer be emitted for as
long as the blood pump is paused.
5.
Call up the
S
TART
/ S
TOP
blood pump view.
6.
In BVAD operating mode, stop both blood pumps; in LVAD or RVAD mode,
stop the corresponding blood pump.
7.
Replace the blood pump. See
EXCOR Sterile Components
instructions for
use.
8.
Perform de-airing in single step mode. Check that the system is completely
de-aired. Do not start the blood pump until the system has been completely
de-aired. Check that connections are air-tight.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...