Configuring and Operating the Driving Unit
en
WARNING
8.10.2 Replacing Driving Tubes - BVAD Right
8.11 Configuring and Updating the Replacement Driving Unit
A configured replacement driving unit must be labeled in such a way
that it is possible to easily allocate it to the correct patient. Otherwise,
switching to the replacement driving unit could result in use of a driving
unit with unsuitable settings. This could lead to insufficient support, no
support, or overloading.
Immediately transfer the changes in the driving parameters to the re-
placement driving unit. Otherwise, the patient will not receive appropri-
ate support if switching to the replacement driving unit becomes
necessary.
8.11.1 Configuring the Replacement Driving Unit
1.
Initial state:
System in
A
UTO
operation. The panel PC is connected to the
driving unit. Driving tube with blue color coding is ready.
2.
Log in with the
E
XPERT
user profile.
3.
Remove the cable wrap.
4.
Remove the flow sensor from the cannula. If alarms are emitted with regard to
flow or a flow sensor not being connected, press
A
UDIO
PAUSED
.
5.
Reduce the rate on the left blood pump if necessary.
6.
Call up the
S
TART
/ S
TOP
BLOOD
PUMPS
view.
7.
Stop the right blood pump.
8.
Replace the right driving tube.
9.
Raise the left blood pump rate back to its original value. Start the right blood
pump.
10.
Check the filling and emptying of the blood pumps and optimize, if necessary.
11.
Mount the flow sensor and cable wrap.
12.
Log off.
1.
Initial state:
The driving unit that is to be configured as a replacement driving
unit is connected to the panel PC. Both devices are on. The configuration file
to be transferred is on the panel PC or on the USB stick.
2.
If the configuration file is found on the USB stick, then insert the USB stick
into the USB port on the panel PC.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...