Configuring and Operating the Driving Unit
en
WARNING
8.8
Postoperative
8.8.1
Monitoring the Patient
Qualified medical personnel logged in as
E
XPERT
or
M
EDICAL
must al-
ways log off of the panel PC before moving away from the patient. Log-
ging off prevents accidental or unauthorized modification of
parameters.
Use intensive-care monitoring to monitor the patient following implantation of a EX-
COR VAD system. Make sure that the user can reliably perceive any potential alarms.
8.8.2
Changing Parameters
The
E
XPERT
and
M
EDICAL
users can change the driving unit parameters. Several pa-
rameters can only be changed after the blood pumps have been paused or the driving
unit has been reset.
Parameters can only be set within a defined range of values. These value ranges de-
pend on the specific components implanted.
1
Lower threshold for the flow alarm
2
Currently selected value
3
Upper threshold for the flow alarm
3.
Set the flow alarm threshold value and use
A
PPLY
to confirm.
Note: If no flow alarm threshold is selected, an alarm will be emitted after 5 minutes
if the rate is 60 bpm or higher.
1
2
3
Parameter
Auto operation
Blood pump(s) stopped
Patient information and
driving unit settings
Parameter settings for the
blood pumps
Blood pump size
Tab. 8-4
Changing parameters
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...