46
1015002x00
Driving Unit
en
5.2.3.4 Removing the Flow Sensor from the Cannula
5.2.3.5 Replacing the Sensor Quick Connect
5.2.3.6 Connecting/Disconnecting the Flow Sensor to/from the Driving Unit
1.
Open the sensor quick connect
(1).
2.
Pinch the sensor locking clip (2)
together at the top.
3.
Remove the sensor locking clip.
4.
Remove the flow sensor from
the cannula.
Fig. 5-25
Removing the flow sensor
1
2
2
1.
Attach the sensor quick connect
(1) to the flow sensor.
2.
Attach the sensor quick connect
(2).
3.
Close (3).
4.
To change the closure side,
rotate the sensor quick connect
(4) or attach it to the other side.
Fig. 5-26
Closing the sensor quick connect
Replace worn or lost sensor quick
connects. Corresponding sensor
quick connects are available for all
sizes of flow sensors.
Fig. 5-27
Sensor quick connect for flow sen-
sors FS06, FS09, FS12
1
2
3
4
06
09
12
1.
Grasp the connector plug under-
neath by the ring.
2.
Align the white markings.
3.
Using light force, press the ring
downwards and turn it in the
direction of the green arrow until
it reaches the stop position.
4.
To release the connection, turn
the ring in the direction of the
white arrow. The plug will
release itself.
Fig. 5-28
Insert the flow sensor
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...