1015002x00
63
Driving Unit
en
5.10.2 Replacing a Driving Unit in LVAD and RVAD Patients
5.10.2.1 Checking the Functionality of an Unconfigured Driving Unit and Panel PC
5.10.2.2 Configuring the Driving Unit from File
5.10.2.3 Checking Configuration and Starting the Driving Unit
1.
Have the panel PC and driving unit ready.
2.
Connect the panel PC to the mains and switch it on.
3.
Connect the driving unit to a power supply. Insert charged batteries. Establish
a mains power connection
4.
The driving unit will conduct a start test. During the start test, the driving tube
connectors must be closed.
5.
Connect the driving unit and panel PC with the USB cable.
6.
The panel PC will log into the driving unit automatically as user
E
XPERT
.
7.
Check the following items on the driving unit:
•
Did the start test run without errors?
•
Are the displays active?
•
Can an audible alarm be heard when the buttons are pressed?
8.
Check the following items on the panel PC:
•
Is a driving unit – panel PC connection in place?
•
Does the user interface respond when the buttons are pressed?
•
Is the user interface bright and easy to read?
9.
If errors occur, see chapter 17: Identifying Errors and Troubleshooting,
page 155.
1.
Initial state:
The configuration file can be found on the panel PC or on a USB
stick.
2.
If the configuration file is on the USB stick, then insert the USB stick into the
USB port on the panel PC.
3.
In the
C
ONFIGURE
DRIVING
UNIT
view, press the
F
ROM
FILE
tab.
4.
The available configuration files will be displayed.
5.
Select the configuration file for the patient, and confirm the selection.
1.
Switch to the Settings view.
2.
Carefully check the settings.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...