56
1015002x00
Driving Unit
en
5.7.1.9 Driving Unit Operating Time in Battery Operation
The operating time of the driving unit depends primarily on the operating mode, the
size of the blood pump, the set rate, and the ambient temperature. The cannulae in
use, the systolic duration, and the patient’s pre- and afterloads all affect operating time
as well.
The following values are guidelines for biventricular operation of 10 mL blood pumps
or an 80 mL/60 mL combination.
BVAD 10 mL blood pumps
Ambient
temperature
Battery oper-
ating time
Operating
time
Emergency
battery
Cannulae 6 mm / 5 mm
Rates 90 bpm / 80 bpm
Systole 30 % / 40 %
Preload 10 mmHg
Afterload 100 mmHg
0 °C
min. 8 h
at least 30
min
25 °C
min. 9 h
40 °C
min. 8:30 h
Operating times are based on new, 100% charged batteries. The operating time
of the emergency battery is measured from the “P7 Emergency battery dis-
charged” notification.
Tab. 5-4
Driving unit operating time in battery operation or emergency battery operation, 10 mL
BVAD 80 mL / 60 mL
Blood pumps
Ambient
temperature
Battery oper-
ating time
Operating
time
Emergency
battery
Cannulae 12 mm / 12 mm
Rates 90 bpm / 80 bpm
Systole 50 %
Preload 10 mmHg
Afterload 100 mmHg
0 °C
min. 5 h
at least 30
min
25 °C
min. 5:15 h
40 °C
min. 5:15 h
Operating times are based on new, 100% charged batteries. The operating time
of the emergency battery is measured from the “P7 Emergency battery dis-
charged” notification.
Tab. 5-5
Driving unit operating time in battery operation or emergency battery operation, 80 mL / 60 mL
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...