Configuring and Operating the Driving Unit
en
WARNING
8.6.4
Setting Rate and Synchronization
Only use rates below 60 bpm during implantation, explantation and
weaning. Do not set rates below 60 bpm on a long-term basis. When
using rates below 60 bpm, monitor blood pump filling and emptying,
and conduct visual checks for deposits and thrombi. At rates below
50 bpm, the driving unit cannot reliably detect losses of pressure, and
may delay emitting alarms.
5.
To stop the blood pumps, the
S
TART
/ S
TOP
button must be
pressed once again. Press the corresponding button to stop
the blood pumps. Only both blood pumps or the right blood
pump can be paused.
The blood pumps can be restarted with
S
TEP
or
A
UTO
.
6.
The start of the pumps is thus concluded. The parameters
must be adjusted subsequently. Check that the blood pumps
are filling and emptying completely.
7.
Continue with chapter 8.6.4: Setting Rate and Synchroniza-
tion, page 104.
1.
The default setting for the
Rate
is 20 bpm. The blood pump
starts with this rate. For regular operation, a value greater than
60 bpm must be selected. To change the rate, press the
R
ATE
button.
2.
Adjust the rate value. Accept the value.
3.
BVAD: In Synchronous mode and Asynchronous mode, the left
and right blood pumps always operate with the same rate. If
the rate for a blood pump is adjusted, the rate for the other
blood pump will be automatically changed.
Differing rates are only possible in
Separate
mode. It is never
possible to select a higher rate for the right blood pump than
for the left blood pump. If the rate for the right blood pump is
increased, the value for the left blood pump will be automati-
cally adjusted. For further information regarding synchroniza-
tion, see Tab. 7-4 , page 82.
4.
To change the synchronization, press the
S
YNCHRONIZATION
button.
5.
Select the type of synchronization and accept the value.
6.
After every rate change, the filling and emptying of the blood
pump must be checked and adjusted if required. Continue with
chapter 8.6.5: Optimizing the Filling and Emptying of the Blood
Pumps, page 105.
7.
After each change in rate, check the
Flow alarm threshold
and change the settings if necessary.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...