User Interface
en
Selection of synchronization
Rate
Number of pump cycles [1/min]
One pump cycle comprises the suctioning and ejec-
tion of blood by the blood pump.
In synchronous and asynchronous modes, setting
the rate in one blood pump simultaneously sets the
rate of the other blood pump.
In separate mode, the rate for each blood pump can
be set separately. Here, the rate for the right side of
the heart can be selected as equal to or smaller than
the rate for the left side of the heart.
Flow alarm threshold
The flow alarm threshold is a freely adjustable
threshold value. If the measurement falls below the
flow alarm threshold, an alarm will be emitted.
Flow
The flow sensor measures the current flow in the
cannula. The measured value is displayed in the
M
ONITORING
window.
The flow cannot be set directly. The flow is automat-
ically calculated based on the rate and on the
selected combination of blood pump and cannulae.
Parameter
Description
Synchronization
Synchronization of the blood pumps can only be
selected in BVAD operating mode. These settings
determine whether the right and left blood pumps
are operated in synchronous, asynchronous, or sep-
arate mode (independently of one another).
Synchronous
mode
The two blood pumps work synchronously. This is
the default setting if the blood pump connected on
the left is larger than the one on the right.
Tab. 7-4
Parameter – Synchronization
Parameter
Description
Tab. 7-3
Parameter – Driving unit blood pumps
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...