1015002x00
65
Driving Unit
en
WARNING
5.11 BVAD Emergency Operation
If one of the driving unit’s pneumatic systems fails in BVAD operating mode, the driv-
ing unit will automatically switch into emergency operation. In emergency operation,
the intact pneumatic unit will supply both blood pumps. Each blood pump is driven for
two pump cycles. After two pump cycles, the switching unit switches the suctioning
and driving pressures to the other blood pump.
At least 45% of the original flow is available during emergency operation. If the driving
unit is in emergency operation, switch to the replacement driving unit as soon as pos-
sible.
In BVAD emergency operation, the blood pumps must be monitored
continuously to ensure complete filling and emptying. Monitor patient
status. If necessary, support the patient using the manual pump. In
emergency operation, the driving unit will not be able to detect insuffi-
cient patient support. Pneumatic leaks cannot be detected with certain-
ty.
9.
In the
B
LOOD
PUMP
SETTINGS
view, select the
F
LOW
ALARM
THRESHOLD
tab. Set
the
Flow alarm threshold
.
10.
Switch off the driving unit that is no longer in use. For shutdown information,
see section 5.6: Switching Off the Driving Unit, page 50. Close all open con-
nectors.
11.
Place the active driving unit onto the transport system.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...