Start Test for the Driving Unit
A driving unit that has not yet been configured will automatically perform a start test
upon starting up. The start test is a routine used to check driving unit functionality. The
start test lasts a maximum of one minute.
Once the start test has completed without errors, the driving unit is ready. After that,
the driving unit can be configured. If an error occurs during the start test, it will be dis-
played.
If an error does not allow for commissioning, this will be shown on the connected panel
PC.
5.5
Driving Unit Self-Test
The self-test is a routine carried out during operation to check the driving unit. It serves
to check back-up components. The self-test also checks whether the driving unit is
due for maintenance.
The self-test is conducted once a day and lasts approximately 30 seconds. The sys-
tem may make different operating sounds during the self-test. Blood pump operation
is not affected.
The self-test time can be configured. Schedule the self-test for a time when the patient
will be awake. That way, the user will be able to react more effectively to messages
like
S1
M
AINTENANCE
DUE
, because messages will not disturb the patient while he or
she is awake.
5.6
Switching Off the Driving Unit
Use the following procedure to switch off the driving unit.
1.
Initial state:
The driving unit is connected to mains power; both batteries are
inserted.
2.
The driving unit is disconnected from all power supplies except one. The
power supplies can be disconnected from the driving unit in any order.
3.
Press both
<buttons>
on the control panel and hold for around five seconds.
Continue holding the buttons until the
Shutdown
symbol appears on both dis-
plays.
4.
Disconnect the driving unit from the remaining power supply within five sec-
onds.
5.
The driving unit is off.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...