User Interface
en
Default Parameter Settings
When the
S
TART
/ S
TOP
button is pressed, a window opens showing buttons to control
the driving unit. The symbols on the buttons show the action to be performed.
Fig. 7-4
Settings – Start-Stop
Asynchronous
mode
The two blood pumps work asynchronously. When
one blood pump is in systole, the other blood pump
is in diastole. Asynchronous mode can be used with
very small patients to minimize intra-thoracic volume
displacement.
Separate mode
Both blood pumps work independently from one
another.
Separate mode
allows for selection of
lower rates for the right side of the heart. Separate
mode is recommended for patients who are supplied
with same-sized blood pumps on the right and left
sides.
Parameter
Default settings
left
right
Synchronization
Synchronous mode
Rate
20 [1/min]
20 [1/min]
Duration systole
50 [%]
50 [%]
Flow alarm threshold
- [l/min]
- [l/min]
Tab. 7-5
Parameter - default settings
1
Step left
With each press of this button, the left blood pump is filled
or emptied.
Pressing 2x equals one pump cycle.
Parameter
Description
Tab. 7-4
Parameter – Synchronization
1
2
4
5
6
3
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...