174
1015002x00
Maintenance, Cleaning and Disinfection
en
Cleaning
Disinfection
18.3.2 Cleaning Agents and Disinfectants, Resources
18.3.3 Cleaning Intervals
Contact areas and functional areas
Contact areas and functional areas on components are areas that may come into di-
rect contact with skin or wounds, or areas that are often touched during operation. The
flow sensor, handles, buttons, keys, plugs, etc. are all areas of frequent contact. They
must be cleaned and disinfected at the intervals specified in Table Tab. 18-5 ,
page 175. They must also be cleaned and disinfected if they have visible contamina-
tion, and whenever they are moved to a different patient.
Periphery
The periphery comprises areas that rarely or never come into direct contact with the
patient, such as the enclosure, the s, or the frame of the caddy. These areas must be
cleaned and disinfected in the event of visible contamination or whenever they are
moved to a different patient.
1.
Moisten a cloth with water and wipe down all exterior surfaces. If
necessary, use a mild, pH-neutral cleaning agent. Larger particles
can first be removed using a soft brush.
2.
If using cleaning agents: Wipe down all surfaces with a damp cloth
to remove any residue. Dry surfaces with a cloth.
1.
Moisten a cloth with disinfectant. Wipe down all exterior surfaces.
Adhere to specified application times.
2.
Moisten a cloth with water. Wipe down all surfaces to remove any
residue. Dry surfaces with a cloth.
Cleaning agent
•
Clean, clear water
•
Surfactant cleaner, pH-neutral
Disinfectant based
on
:
•
ethanol
•
1-propanol
•
2-propanol
Use disinfectants without other additives if
possible.
Resources
•
Clean, soft, lint-free cloth
•
Soft brush
Tab. 18-4
Cleaning Agents and Disinfectants, Resources
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...