23.1 Safety Information
Use of EXCOR Active with accessories or components (e.g., power
supply cables) that have not been specified and supplied by the
EXCOR Active system manufacturers could create increased electro-
magnetic disturbances or reduce the electromagnetic immunity of
EXCOR Active. This may lead to malfunctions.
Ensure that a distance of at least 30 cm is maintained between
EXCOR Active and any active mobile phones or portable HF commu-
nication equipment (including accessories such as antenna cables or
external antennas). Otherwise, electromagnetic disturbances may oc-
cur, and the functionality of EXCOR Active cannot be guaranteed.
Additional devices connected to medical electrical equipment must
comply with IEC or ISO standards (e.g. IEC 60950 or IEC 62368 for
data processing equipment). In addition, all configurations must meet
the requirements for medical electrical systems (see section 16 of the
latest valid version of IEC 60601-1). Anyone who connects additional
devices to medical electrical equipment configures a medical system
and is therefore responsible for ensuring that the system meets the re-
quirements for medical electrical systems. In case of doubt, please
contact Berlin Heart.
23.2 Major Performance Characteristics
The following essential performance characteristics have been tested, and are valid
even under the influence of electromagnetic disturbances:
•
Automatic start of a configured driving unit when connected to a power supply
•
Blood pump operation using the set parameters
•
Continuous monitoring of driving unit pressures and blood flow in order to detect
changes in patient care and emit any necessary alarms.
•
Audible and visible emission of high-priority alarms
The following parameters can be set on EXCOR Active to the specified degrees of ac-
curacy (for right and left in BVAD mode, if applicable):
•
The
Rate
can be set within a range of 5 bpm to 130 bpm. The parameter does
not deviate from the set value by more than ± 5%.
•
Duration systole
can be set within a range of 30% to 60%. The parameter does
not deviate from the set value by more than ± 5%.
•
The
Average driving pressure
does not deviate from the set value by more than
± 10 mmHg.
•
The
Piston stroke
does not deviate from the set value by more than ± 2 mm.
•
The
Flow
does not deviate from the set value by more than ± 10%.
•
Synchronization
does not deviate by more than ± 8%.
Even under the influence of interference, the following apply:
•
In normal operation, no false positive alarms of low, medium, or high priority oc-
cur. Electromagnetic radiation has no effect on the driving unit’s capability of emit-
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...