Technical Alarm Messages at the TD60
Troubleshooting
14 - 8
TMS60 Operator’s Manual
NIBP
NIBP
Error
High
A
1.
An error occurred
to the NIBP
module.
2.
There is a problem
with
communication
between the TD60
and BP10.
Restart the BP10.
NIBP Cuff
Loose
Low
C
1.
The NIBP cuff is
not properly
connected.
2.
There is a leak in
the airway.
1.
Check the
patient's con-
dition and
verify patient
category.
2.
Replace with
an appropri-
ate cuff and
connect it
correctly.
NIBP
Airway
Error
Low
C
An airway error occurs.
Check the airway.
NIBP
Weak
Signal
Low
C
The patient’s pulse is
weak or the cuff is loose.
Check the patient’s
condition and
change the cuff
application site. If
the error persists,
replace the cuff.
NIBP
Over-
range
Low
C
The measured NIBP value
is not within the specified
range.
Contact your service
personnel.
NIBP
Exces-
sive
Motion
Low
C
Patient’s arm moves too
much.
Check the patient’s
condition and
reduce the patient
motion.
NIBP Cuff
Over-
pressure
Low
C
The NIBP airway may be
occluded.
Check the airway
and measure again.
Measurement
Alarm
message
Alarm
level
Alarm
Indication
Possible cause
Solution
*
The asterisk (*) means the alarm level is configurable.
**
XX is the ECG lead name.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...