Monitoring with the TMS60 at the CMS
Monitoring at the CMS (Version Earlier than 03.00)
TMS60 Operator’s Manual
12 - 9
12.2.2
ECG Monitoring
At the CMS, you can view and change the heart rate (HR), QT, ST, and arrhythmia settings
in the
ECG
tab of the
Parameter Setup
menu.
For details about the
Parameter Setup
menu, See the
BeneVision Central Monitoring
System Operator’s Manual
.
At the CMS, you can view and change the ECG settings in the
ECG
tab. To enter the ECG
tab, follow this procedure:
1.
Select the ECG digital area or waveform area to enter the
Parameter Setup
menu.
2.
Select the
ECG
tab.
12.2.2.1
HR Settings
The following table lists the HR settings in the
HR
section of the
ECG
tab.
Options or Buttons
Description
Settings*
HR Activation State
Configures whether or not to enable the
HR alarm.
Alarm Settings" on page 12 - 3 for
details.
HR Alarm Priority
Configures the HR alarm levels.
HR High Limit (bpm)
Configures the HR high alarm limit.
HR Low Limit (bpm)
Configures the HR low alarm limit.
Sweep Speed
Configures the ECG wave speed.
6.25 mm/s, 12.5 mm/s,
25 mm/s
,
50 mm/s
HR Source
Configures the HR source.
ECG
, SpO
2
, Auto, Both
HR Record on Alarm
Select whether or not to activate the
option to direct the CMS to send the HR
alarm data to the configured recorder.
On,
Off
Waveform Setup
Select to display the
Waveform Setup
menu.
for details.
Other Settings
Select to display the
Other Settings
menu.
for details.
*
The factory default settings are in bold.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...