Monitoring with the TMS60 at the CMS
Monitoring at the CMS (Version 03.00 or Above)
TMS60 Operator’s Manual
12 - 37
12.3.2.3
Changing Pacer Settings
To change pacer related settings, follow this procedure:
1.
From the
ECG
tab, select the
Pacer
section.
2.
Set the desired items.
12.3.2.4
Adjusting the Minimum QRS Detection Threshold
To avoid false asystole alarms when the R wave amplitude is low and missed asystole
alarms during ventricular standstill (tall P waves, but no QRS), a means to manually
adjust the minimum QRS detection threshold is provided.
To adjust the QRS detection threshold, follow this procedure:
1.
From the
ECG
tab, select the
Setup
section.
2.
From the
ECG
tab, select the
QRS Threshold
section.
3.
Select the arrow buttons to adjust the QRS threshold. Selecting
Default
resets the
QRS threshold to the default value (0.16 mV).
Options or Buttons
Description
Default Setting
Paced
The option is not configurable at the CMS.
The paced status can be set at the TD60
only. See
for details.
None
Pacer Reject
Configures whether or not to reject the
pace pulses.
■
On: the pace pulses are not
counted as extra QRS complexes.
■
Off: the pace pulses are not
rejected.
Off
This option is only available when
Paced
is set to
Yes
at the TD60.
CAUTION
•
The setting of QRS threshold can affect the sensitivity of arrhythmia, ST,
QT/QTc detection, and heart rate calculation.
•
If QRS amplitude is low, the monitor might not be able to calculate heart
rate and false asystole may occur.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...