Preparation for Monitoring ECG
Monitoring ECG
7 - 10
TMS60 Operator’s Manual
7.3.6
Setting the Paced Status
It is important to correctly set the patient’s paced status before you start monitoring
ECG.
To set the paced status, follow this procedure:
1.
Enter the ECG menu in either of the following ways:
◆
On the main screen, tap the HR digital area or ECG waveform area to enter
the
ECG
menu.
◆
Press the
button to enter the main menu, tap
Parameter Setup
, and
then tap
ECG
.
2.
In the Paced section, tap
Paced
and select the correct paced status.
◆
When
Paced
is set to
Yes
at the TD60, if pace pulses are detected, the
symbol displays in the waveform area of the Central Monitoring System’s
screen, and the pace pulse marks will display on the ECG waveform both at
the TD60 and CMS.
◆
When
Paced
is set to
No
or not specified at the TD60, the
symbol dis-
plays in the waveform area of the Central Monitoring System’s screen.
WARNING
•
For paced patients, you must set Paced to Yes. If it is incorrectly set to No,
the CMS could mistake a pace pulse for a QRS and fail to alarm when the
ECG signal is too weak.
•
On ventricular paced patients, episodes of ventricular tachycardia may-
not always be detected.
•
Do not rely entirely on rate meter alarms when monitoring patients with
pacemakers. Always keep these patients under close surveillance.
•
For non-paced patients, you must set Paced to No. The pacer pulses may
be counted as QRS complexes, hence leading to wrong HR readings or
failure to diagnose certain arrhythmia symptoms. Be sure to keep a
close eye on patient’s with pacemaker devices.
•
The radio frequency energy of the transmitter or other radio frequency
sources, when used in close proximity to a pacemaker, may interfere
with pacemaker performance. Internal pacemakers are less vulnerable
than external pacemakers due to the shielding effects of the body. How-
ever, caution should be exercised when monitoring any paced patient.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...