Monitoring at the CMS (Version Earlier than 03.00)
Monitoring with the TMS60 at the CMS
12 - 8
TMS60 Operator’s Manual
12.2.1.3
Arrhythmia Threshold Settings
When an arrhythmia violates its threshold, an alarm is triggered. When HR is less than 30
bpm, it is recommended to set asystole delay to 10 s.
Arrhythmia event
Range or Option
Default
Step
Unit of
measure
PVCs High
1 to 100
10
1
minute
Pauses/min
1 to 15
8
1
None
Asystole Delay
3 to 10
4
1
second
Tachy High
60 to 300
Adult: 120
Pediatric: 160
5
bpm
Brady Low
15 to 120
Adult: 50
Pediatric: 75
5
bpm
Extreme Tachy
60 to 300
Adult: 160
Pediatric: 180
5
bpm
Extreme Brady
15 to 120
Adult: 35
Pediatric: 50
5
bpm
Multif PVCs Window
3 to 31
15
1
/min
V-Tach Rate
100 to 200
130
5
bpm
V-Tach PVCs
3 to 99
6
1
beat
Pause Threshold
1.5, 2, and 2.5
2
None
second
V-Brady Rate
15 to 60
40
5
bpm
V-Brady PVCs
3 to 99
5
1
beat
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...