Monitoring at the CMS (Version 03.00 or Above)
Monitoring with the TMS60 at the CMS
12 - 34
TMS60 Operator’s Manual
In addition, the activation state of some arrhythmias can be set as a whole with the fol-
lowing buttons that are at the bottom of the
Arrhythmia Alarms
tab.
Missed Beats
On,
Off
High, Med, Low,
Prompt
On,
Off
Multiform PVC
High,
Med
, Low, Prompt
Nonsus V-Tach
Pause
High, Med,
Low
, Prompt
A-Fib
High, Med, Low,
Prompt
Irr Rhythm
Low,
Prompt
Pauses/min
High,
Med
, Low, Message
Name
Activation State
Alarm Priority
Alarm
Outputs
Notes
The factory default settings are in bold.
NOTE
•
When Paced is set to Yes, the Missed Beat (MIS) alarm is reported as the
Pacer Not Capture
(PNC) or
Pacer Not Pacing
(PNP) alarm.
•
The priority of lethal arrhythmia alarms is always high. It is unchange-
able.
Button
Description
Lethals Only
Sets the lethal arrhythmia alarms to on and all non-lethal arrhythmia alarms to off.
All On
Sets all arrhythmia alarms to on.
All Off
Sets all arrhythmia alarms to off.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...