Monitoring with the TMS60 at the CMS
Monitoring at the CMS (Version Earlier than 03.00)
TMS60 Operator’s Manual
12 - 23
12.2.5.1
Measurement Limitations
ƾ
Since the arrhythmia detection algorithm sensitivity and specificity is less than
100%, sometimes there may be some false arrhythmias detected and also some
true arrhythmia events may not be detected. This is especially true when the
signal is noisy.
ƾ
The ECG size and QRS width settings affect arrhythmia detection and heart rate
calculation sensitivity.
ƾ
If QRS amplitude is low, the monitor might not be able to calculate HR and false
asystole may occur.
ƾ
At the algorithm learning phase, arrhythmia detection may not be available.
Therefore, closely monitor the patient’s condition during the learning phase after
the learning phase to allow the algorithm to reach optimal detection
performance.
12.2.5.2
Understanding the Arrhythmia Events
Arrhythmia message
Description
Category
Asystole
No QRS detected within the set time threshold in absence of ven-
tricular fibrillation or chaotic signal.
Lethal
arrhythmia
V-Fib/V-Tach
A fibrillatory wave for 6 consecutive seconds.
A dominant rhythm of adjacent Vs and a HR > the V-Tach HR limit.
V-Tach
The consecutive PVCs ≥ V-Tach PVCs limit, and the HR ≥ the V-Tach
Rate limit.
Vent Brady
The consecutive PVCs ≥ the V-Brady threshold and the ventricular
HR < the V-Brady Rate threshold.
Extreme Tachy
The heart rate is equal to or greater than the extreme tachycardia
limit.
Extreme Brady
The heart rate is equal to or less than the extreme bradycardia limit.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...