Monitoring with the TMS60 at the CMS
Monitoring at the CMS (Version 03.00 or Above)
TMS60 Operator’s Manual
12 - 31
Parameters
Activati
on
State
Alarm High
Limit
Alarm Low
Limit
Alarm
Priority
Alarm Outputs
HR/PR
Unit of mea-
sure: bpm
Invalid data:
“--”
O
n
, Off
For adult:
(Low limit + 2)
to 300
Default: 120
For pediatric:
(Low limit + 2)
to 350)
Default: 160
For adult:
15 to (high
limit - 2)
Default: 50
For pediat-
ric:
15 to (high
limit - 2)
Default: 75
High,
Med
On,
Off
ST Single
Unit of mea-
sure: mV
Invalid data:
“--”
On,
Off
For adult and
pediatric:
(Low limit +
0.2) to 2.00
Default: 0.1
For adult
and pediat-
ric:
-2.00 to
(high limit –
0.2)
Default: -0.1
High,
Med
,
Low
ST Dual
Unit of mea-
sure: mV
Invalid data:
“--”
QTc
Unit of mea-
sure: ms
Invalid data:
“--”
For adult:
200 to 800
Default: 500
For pediatric:
200 to 800
Default: 480
For adult
and pediat-
ric: None
∆
QTc*
Unit of mea-
sure: ms
Invalid data:
“--”
For adult and
pediatric:
30 to 200
Default: 60
For adult
and pediat-
ric: None
RR
Unit of mea-
sure: rpm
Invalid data:
“--”
On
,
Off
For adult and
pediatric:
10 to 100
Default: 30
For adult
and pediat-
ric: 0 to 55
Default: 8
Notes:
The factory default settings are in bold.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...