If both V-Tach PVCs and V-Tach Rate are greater than or equal to the limits, a V-Tach alarm
is generated.
ƾ
If consecutive PVCs is lower than the V-Tach PVCs limit (6) but greater than 2, and
the Vent rate is greater or equal to the V-Tach Rate limit (130), a Nonsus V-Tach
alarm is generated.
ƾ
If consecutive PVCs is greater than or equal to the V-Brady PVCs limit (5), and the
Vent Rate is lower than the V-Brady limit (40), a Vent Brady alarm is generated.
ƾ
If both the V-Brady PVCs and V-Brady Rate are lower than the limits, but V-Brady
PVCs is greater than 2, a Run PVCs alarm is generated.
ƾ
If the V-Brady PVCs and V-Brady Rate are greater than or equal to limits, but the
Vent rate is is lower than V-Tach Rate (130), a Vent Rhythm alarm is generated.
12.3.5.6
Arrhythmia Display
When
Arrhythmia
is selected to display in the
Tile Layout
, the arrhythmia digital area
displays:
1.
Parameter label
2.
Measurement unit
3.
Arrhythmia alarm off symbol
4.
PVCs per minute label
5.
Pauses per minute label
6.
PVCs per minute value
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...