Monitoring at the CMS (Version Earlier than 03.00)
Monitoring with the TMS60 at the CMS
12 - 12
TMS60 Operator’s Manual
3. Select
Exit
to save the settings and close the
Analysis Lead Setup
menu.
Other Settings
In the
HR
section of the
ECG
tab, select
Other Settings
. The
Other Settings
menu dis-
plays.
The following table lists all settings in the
Other Settings
menu.
12.2.2.3
ECG Display
The ECG digital area displays:
1.
Parameter label
2.
HR value
3.
High HR alarm limit
4.
Low HR alarm limit
5.
Pauses threshold
6.
Pauses per minute value
7.
Pauses per minute label
8.
PVCs threshold
9.
PVCs per minute value
10.
PVCs per minute label
Options
Description
Settings*
Paced
The option is unavailable at the CMS. The paced status can be set at the TD60 only.
Refer to
"Setting the Paced Status" on page 7 - 10
for details.
Pacer Reject
Configures whether or not to reject the pace
pulses.
■
On
: the pace pulses are not counted as
extra QRS complexes.
■
Off
: the pace pulses are not rejected.
On,
Off
This
Pacer Reject
option is only available when
Paced
is set to
Yes
at the TD60.
Filter
Configures the ECG filter in all operating
modes.
■
Monitor
: use under normal
measurement conditions.
■
ST
: use when the ST monitoring is
applied.
Monitor
, ST
*
The factory default settings are in bold.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...