BeneVision TM80 Telemetry Monitor Operator’s Manual
15 - 9
Measurement
Alarm
message
Default
Alarm
priority
Alarm
Indication
Possible cause
Solution
ECG
ECG Lead
Off
Low*
B
The electrode
has become
detached from
the patient or
the lead wire has
become
disconnected
from the adapter
cable.
Check the
connections of
the electrodes
and leadwires.
ECG XX**
Lead Off
Low*
B
ECG
Module
Error
High
A
An error
occurred to the
ECG module.
There is a
problem with
the
communications
between the
module and the
telemetry
monitor.
Restart the
telemetry
monitor. If the
problem persists,
contact your
service personnel.
*
The asterisk (*) means the alarm priority is configurable.
**
XX is the ECG lead name.
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...