10 - 2
BeneVision TM80 Telemetry Monitor Operator’s Manual
6.
Tap anywhere in the NIBP parameter area on the telemetry monitor’s main screen to
access the
NIBP
menu.
7.
Tap
Start
to start NIBP measurement. Verify that the NIBP measurement results dis-
played at the BP10 are consistent with that displayed on the main screen of the
telemetry monitor and are the data for the same patient.
WARNING
•
Before pairing a BP10 with a new telemetry monitor, unpair it with the current
telemetry monitor first.
•
Do not pair a telemetry monitor with an undesired BP10. If a telemetry monitor
is paired with an undesired BP10, the patient category may be not applicable
for the patient and the measured NIBP data may be incorrect.
T.
NOTE
•
Make sure that you have selected the correct BP10 to pair it with the telemetry
monitor.
•
If you pair a telemetry monitor with a BP10 when NIBP measurement mode is
set to ABPM at BP10, the system responses after successful pairing are slightly
different from that when the NIBP measurement mode is set to Manual, Auto, or
Sequence. For the differences, see
10.3.3 System Responses after Successful
10.2.2 System Responses after Successful Pairing
Once a telemetry monitor is paired successfully with a BP10, the following system
responses occur:
■
The
icon will be changed to
in the top-right corner of the telemetry
monitor’s screen. Besides, the
appears in the top-right corner of the BP10.
■
The status of the BP10 which is paired successfully with the telemetry monitor is
displayed as
Connected
. Besides, the “Pairing Successful” displays in the BP10’s
message area for about 10 seconds.
■
When NIBP measurement mode is set to
ABPM
at BP10 and the BP10 is paired
successfully with the telemetry monitor, the wireless seup confirmation message
“Select Yes to continue Patient in BP10. Select No to continue Patient in Telemetry.”
will display on the main screen of the telemetry monitor.
◆
Yes
: Patient-related settings comes from BP10 and ongoing NIBP measurement
for current patient will not stop.
◆
No
: Patient-related settings at BP10 comes from the telemetry monitor.
Besides, NIBP measurement for the patient monitored by BP10 will stop and
the data history related to this patient will be cleared at BP10. When a BP10 in
ABPM mode is not unpaired with a telemetry monitor and needs to be paired
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...