When the telemetry monitor is connected to the CMS, you can adjust ST points, enter the
ST graphic window and the ST view at the CMS. For more information, refer to
7.8.1 ST Safety Information
WARNING
•
ST values may be affected by such factors as some drugs or metabolic and
conduction disturbances.
•
When the telemetry monitor is connected to the CMS, the CMS provides
information regarding changes in ST deviation levels. The clinical significance
of the ST level change information should be determined by a physician.
7.8.2
Enabling ST Monitoring
The ST monitoring function is disabled by default. Before you start ST monitoring, enable
the ST function. To do so, follow this procedure:
1.
On the main screen, tap the HR digital area or ECG waveform area to enter the
ECG
menu.
2.
In the
Parameter Setup
section, tap
ST
.
3.
Switch on
ST Analysis
.
Reliable ST monitoring can hardly be ensured if:
■
You are unable to get a lead that is not noisy.
■
Arrhythmias, such as atrial fibrillation or atrial flutter, which may cause an irregular
baseline.
■
The patient is continuously ventricularly paced.
■
The patient has left bundle branch block.
In these cases, you may consider switching off ST monitoring.
7.8.3 Displaying ST Numerics
When
ST Analysis
is switched on, ST digital area is displayed on the screen. The location of
the ST digital area is dependent on the display configurations. For how to set the location
of the ST digital area, refer to
.
Numerics displayed in the ST digital area are different according to the lead type:
■
When the 3-lead ECG leadwires are used, the ST digital area is not displayed. A ST
value is displayed in the HR digital area.
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...