BeneVision TM80 Telemetry Monitor Operator’s Manual
15 - 7
SpO
2
SpO
2
High
Med*
SpO
2
value has
risen above the
high alarm limit or
fallen below the
low alarm limit.
Check the patient’s
condition and
make sure that the
patient category
and alarm limit
settings are
correct.
SpO
2
Low
SpO
2
Desat
High
The SpO
2
value has
fallen below the
desaturation alarm
limit.
PR High
Med*
PR value has risen
above the high
alarm limit or fallen
below the low
alarm limit.
PR Low
No Pulse
High
The pulse signal
was so weak that
the monitor cannot
perform pulse
analysis.
Check the patient’s
condition, SpO
2
sensor, and
measurement site.
NIBP
NIBP-S High
Med*
The NIBP systolic
pressure has risen
above the high
alarm limit or fallen
below the low
alarm limit.
Check the patient’s
condition and
check if the alarm
limit settings and
patient category
are correct.
NIBP-S Low
NIBP-M High
The NIBP mean
pressure has risen
above the high
alarm limit or fallen
below the low
alarm limit.
NIBP-M Low
NIBP-D High
The NIBP diastolic
pressure has risen
above the high
alarm limit or fallen
below the low
alarm limit.
NIBP-D Low
Measurement
Alarm
message
Default
Alarm
priority
Possible cause
Solution
*
The asterisk (*) means the alarm priority is configurable.
** XX is the ECG lead name
.
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...