7 - 14
BeneVision TM80 Telemetry Monitor Operator’s Manual
NOTE
•
When Paced is set to Yes, the Markers and Pacer Rejection options can be
available.
7.4.6 Configuring the ECG Waveform Size
The
Waveform Size
field of the
ECG
menu lists all available leads. You can select the
desired ECG lead to set the waveform size. For details about the waveform size setting, refer
to
7.4.4 Configuring ECG Waveforms
.
7.4.7 Configuring ECG Alarm Settings
To configure ECG alarm settings, follow this procedure:
1.
In the
Alarms
section of the
ECG
menu, tap
ECG Alarm Setup
.The
Alarm Limits
menu is displayed.
2.
Select the options described in the following table.
7.4.8 Configuring the Notch Filter
Notch filter filters out AC line noise from the ECG waveform. Refer to
for details.
Options
Description
Settings*
HR/PR
Configures whether to
trigger a HR/PR high or low
alarm.
On
, Off
Alarm limit range:
16 bpm to 299 bpm; the step is 1bpm.
The default alarm high limit is
120
for adult
and is
160
for pediatric.
The default alarm low limit is
50
for adult
and is
75
for pediatric.
The alarm high limit should be at least 2
bpm greater than the alarm low limit.
Alarm priority:
Med
, High
Note:
1. When you change the high limit of HR/PR, the limit of Tachy will be automatically
changed, and vice versa.
2. When you change the low limit of HR/PR, the limit of Brady will be automatically
changed, and vice versa.
3. When you change the alarm priority of HR/PR, the alarm priority of Tachy/Brady
will be automatically changed, and vice versa.
*
The factory default settings are in bold.
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...