17.6 Verifying the ECG
The ECG signal may be inaccurate due to hardware or software problems. As a result, the
ECG wave amplitude becomes greater or smaller. To verify the ECG waveform amplitude:
1.
In the main menu, tap
Parameter Setup
.
2.
Tap
ECG
and set
Filter
to
Monitor
.
3.
In the main menu, tap
Maintenance
.
4.
Input the maintenance passcode and tap
Accept
.
5.
In the
Maintenance
menu, tap
General
.
6.
Enable
Calibrate ECG
.
A square wave appears on the screen and the message
ECG Calibrating
is displayed.
7.
Compare the amplitude of the square wave with the wave scale. The difference
should be within 5%. If the difference exceeds 5%, contact your service personnel.
8.
After completing the verification, disable
Calibrate ECG
.
17.7 NIBP Accuracy Test
The NIBP accuracy test is required at least once a year or when you doubt the measured
NIBP result. Contact your service personnel to perform NIBP accuracy test.
17.8 NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at
least once a year or when you doubt the measured NIBP result. Contact your service
personnel to perform NIBP leakage test.
17.9 Nurse Call Test
The nurse call test checks whether this functionality can work properly. Contact your
service personnel to perform the nurse call test.
17.10 Electrical Safety Tests
The electrical safety tests aim to detect abnormalities that, if undetected, could prove
dangerous to either the patient or the operator. Contact your service personnel to perform
the electrical safety tests.
17.11 Battery Check
For details about the battery charge check and maintenance, refer to
Rechargeable Lithium-ion Battery
.
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...