BeneVision TM80 Telemetry Monitor Operator’s Manual
13 - 9
13.7 Arrhythmia Monitoring
13.7.1 Configuring the PVC-Related Alarm Threshold
PVC-related alarms are detected on the basis of the current ventricular heart rate and the
number of consecutive PVCs.
To configure the threshold of PVC-related alarm threshold, follow this procedure:
1.
On the ViewBed screen of the CMS, select the ECG digital area or waveform area to
enter the
ECG
menu.
2.
Select the
Arrhythmia
tab.
3.
Select the
More
Threshold
section.
4.
Set the desired items using the right or left arrow.
The following figure illustrates the conditions under which PVC alarms will be generated if
V-Tach PVCs
is set to 6,
V-Tach Rate
is set to 130,
V-Brady PVCs
is set to 5, and
V-Brady
Rate
is set to 40.
Display/Hide Baseline
Displays or hides the reference
baseline.
None
Display/Hide Marker
Displays or hides the position of
ISO point, J point and ST point.
Buttons
Description
Settings*
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...