15 - 6
BeneVision TM80 Telemetry Monitor Operator’s Manual
ECG
QTc High
Med
*
The QT value has
risen above the
high alarm limit.
Check the
patient’s
condition and
check if the
patient category
and alarm limit
settings are
correct.
ΔQTc High
Med
*
Resp
RR High
Med*
RR value has risen
above the high
alarm limit or fallen
below the low
alarm limit.
Check the patient’s
condition and
make sure that the
patient category
and alarm limit
settings are
correct.
RR Low
No Chest
Movement
High
The respiration
signal was so weak
that the monitor
cannot perform
respiration analysis.
Check the patient’s
condition, module
and patient
connections.
Resp Artifact
The patient’s
heartbeat has
interfered with his
respiration.
Measurement
Alarm
message
Default
Alarm
priority
Possible cause
Solution
*
The asterisk (*) means the alarm priority is configurable.
** XX is the ECG lead name
.
.
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...