13 - 2
BeneVision TM80 Telemetry Monitor Operator’s Manual
WARNING
•
It is not recommended to assign a telemetry monitor to a department in the
Configuration tab of the CMS.
13.2 Network Safety Information
CAUTION
•
Wireless network designing, deploying, debugging, and maintenance should
be executed by Mindray service personnel or authorized technicians.
•
Always set the wireless network according to local wireless regulations.
•
Keep network authentication information, for example password, safe,
protecting the network from being accessed by unauthorized users.
•
Do not connect non-medical devices to the monitor network.
•
If wireless network signal is poor, there may be a risk of CMS data loss.
•
RF interference may result in wireless network disconnection.
•
Ensure that the TM80’s IP address setting is correct. Changing the network
settings may result in network disconnection. Contact your service personnel if
you have any problems on setting the IP address.
•
Disconnecting from the network may result in CMS data loss and function
failure. Check the patient in case of network disconnection and solve the
network problem as soon as possible.
13.3 Adjusting the Minimum QRS Detection Threshold
To avoid false asystole alarms when the R wave amplitude is low and missed asystole
alarms during ventricular standstill (tall P waves, but no QRS), a means to manually adjust
the minimum QRS detection threshold is provided.
To adjust the QRS detection threshold, follow this procedure:
1.
On the ViewBed screen of the CMS, select the ECG digital area or waveform area to
enter the
ECG
menu.
2.
From the
ECG
tab, select the
QRS Threshold
section.
3.
Select the arrow buttons to adjust the QRS threshold. Selecting
Defaults
resets the
QRS threshold to the default value (0.16 mV).
CAUTION
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...