6416h GB/2021-01
47 / 86
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
3.7
Mounting/removal of the TRISON Twist
If the TRISON Twist Si or Xi is not needed, it can be easily removed without tools and
stored in a di
ff
erent clean and dry location. It can be reattached just as easily.
Important:
The plug on the TRISON Twist must not be submerged!
If the plug comes into contact with liquids it must be blown out with compressed air and
allowed to dry completely.
Removal:
Before the TRISON Twist Si or Xi is removed, withdraw all instruments.
Undo all connections between the TRISON Base and TRISON Twist (push-pull plug
and hose connections).
Hold the plug in your hand and protect it from liquid.
Hold the TRISON Twist by its handle and pull directly upwards.
Then store it in a clean, dry location.
Attachment:
Hold the TRISON Twist Si or Xi by its handle. Using its guide, slide it straight
downward into the head of the TRISON Lift, until the dead stop is reached and it
clicks into place.
Then the TRISON Twist can be loaded with the instruments.
View: