1.6.2.3 Possible side e
ff
ects/limitations
• Processes such as chemical disinfection can be accelerated (time-shortened) in
TRISON ultrasonic baths. Ultrasound alone does not disinfect.
• Cavitation erosion can cause surfaces to be mechanically corroded and their
coatings to dissolve.
• In case of steam formation, the TRISON ultrasonic baths must be operated with a lid.
• Ultrasound warms up the bath liquid even without additional heating.
1.6.3 User
group
TRISON ultrasonic baths are intended for commercial use, e.g. in a treatment unit for
medical devices (AEMP). The ultrasonic baths must be used by medical or equivalent
skilled personnel.
1.7
CE conformity
The ultrasonic bath is declared as a medical device and satis
fi
es the CE-marked
criteria for the European directives:
– Medical Device Directive
– Low-Voltage Directive
– Electromagnetic Compatibility Directive
– RoHS Directive
in their currently valid versions.
A declaration of conformity can be requested from the manufacturer by providing the
serial number.