English–DPI 610/615 IS Instruction Manual | 33
Data Log Operations
6.4.7 Running Procedure Files
To run a procedure (DPI 615 instruments only), make sure that the instrument is set to Store
mode, Data Log (see Section
7.2), and proceed as follows:
After selecting F1, proceed by entering the User ID and Serial Number and then select F1
(Continue) and follow the on-screen procedural instructions:
When the test procedure for a given UUT has been completed, the result of running the first test
is stored as an AS FOUND file. This file cannot be overwritten. Any subsequent tests on the
device are stored as an AS LEFT file that is overwritten each time the procedure on this device
is run.
When recalling the results of a procedure, the choice of AS FOUND or AS LEFT is provided (refer
to Section
6.4.8 Recalling Data Files
Data or results files generated by running procedures (DPI 615 instruments only) are stored in
the instrument’s data log directory. To recall a data file to the display, make sure that
DATA LOG
is selected from the
SETUP
menu and proceed as follows:
Use the cursor keys to select either the AS FOUND option or the AS LEFT option for display. AS
FOUND is the result of the first run of a procedure and AS LEFT is the result of the last time the
procedure was run.
Procedure data files can be displayed either as text (stored screens) or in graphical form. To
display as text, select the
TEXT
option (F1) from the directory and proceed as follows from the
A
Setup and Log Point 1
B
Point 1 stored
C
Setup and Log Point 2 etc.
MEMORY USED
FILE 1 of 9
DATE: 04/11/99 TIME 10:18
ERASE
SELECT FILE
RUN
PROCEDURE
REVIEW
LOGGED FILE
FEEDPRESS11
DEVICE UNDER TEST
I/P LOW:
I/P HIGH:
O/P LOW:
O/P HIGH:
MAX PRESS:
POINTS: 5
UP/DOWN
MAX. ERROR: 0.1% SPAN
0.04 %span PASS
0.04 %span PASS
LOG
NOW
LOG
NOW
DATA POINT LOGGED
GO TO 0.000 bar
GO TO 0.000 bar
GO TO 5.000 bar
MEMORY USED
FILE 1 of 9
DATE: 04/11/99 TIME 10:18
ERASE
SELECT FILE
RUN
PROCEDURE
REVIEW
LOGGED FILE
TEXT
GRAPH
DIRECTORY
AS FOUND
AS LEFT
FEEDPRESS11
DEVICE UNDER TEST
I/P LOW:
I/P HIGH:
O/P LOW:
O/P HIGH:
MAX PRESS:
POINTS: 5
UP/DOWN
MAX. ERROR: 0.1% SPAN
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......