42 | DPI 610/615 IS Instruction Manual–English
Chapter
8.
Calibration
3. Switch the instrument on, enter
CALIBRATION
mode and select
INT
RANGES
from the
CALIBRATION
menu and then
PRESSURE
INT
.
4. Apply zero pressure and store the zero point:
5. Close the vent valve, apply full-scale pressure and store the full-scale point:
6. Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to quit
CALIBRATION
and
SETUP
modes.
7. Check calibration by applying test pressures in Table
40.
8.10 Voltage Input Range (5 V)
Use the following procedure for calibrating the internal 5 V range:
1. Switch the instrument on, enter
CALIBRATION
mode and select
INT
RANGES
from the
CALIBRATION
menu shown in Section
PRESSURE INT
CURRENT IN
VOLTAGE IN
CALIBRATION
SELECT OPTION
CURRENT OUT
TEMPERATURE
CALIBRATING PRESSURE INT
RANGE 20.000 bar g
Apply nominal zero value
Press ENTER key when
applied value is stable
CALIBRATING PRESSURE INT
RANGE 20.000 bar g
Enter Applied Value:
CHANGE
VALUE
RUN
SET VALUE
CALIBRATING PRESSURE INT
RANGE 20.000 bar g
Store zero point?
YES
NO
CALIBRATING PRESSURE INT
RANGE 20.000 bar g
Apply nominal FS value
Press ENTER key when
applied value is stable
CALIBRATING PRESSURE INT
RANGE 20.000 bar g
Enter Applied Value:
CHANGE
VALUE
RUN
CHANGE
VALUE
CALIBRATING PRESSURE INT
RANGE 20.000 bar g
Enter Applied Value:
CHANGE VALUE
CALIBRATING PRESSURE INT
RANGE 20.000 bar g
Store FS point?
YES
NO
Calibration Complete
Press ENTER to continue
CALIBRATING PRESSURE INT
RANGE 20.000 bar g
RESULTS
Applied
Measured
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......