46 | DPI 610/615 IS Instruction Manual–English
Chapter
8.
Calibration
8.12 Current Input Range (55 mA)
Use the following procedure for calibrating the current input range:
1. Switch the instrument on, enter
CALIBRATION
mode and select
INT
RANGES
from the
CALIBRATION
menu as shown in Section
8.8.
2. Select
CURRENT
IN
range for calibration:
3. Open circuit the
mA
IN
terminals and enter the zero point:
Table
8
-
3: Electrical Voltage Input Calibration Tolerances
Voltage Calibration 30 V Range
Verification Tolerances
Voltage Calibration 5 V Range
Verification Tolerances
Applied Voltage
Permissible Deviation
Applied Voltage
Permissible Deviation
-30
± 0.05% rdg, ± 0.004% FS
-5
± 0.05% rdg, ± 0.004% FS
-24
± 0.05% rdg, ± 0.004% FS
-4
± 0.05% rdg, ± 0.004% FS
-18
± 0.05% rdg, ± 0.004% FS
-3
± 0.05% rdg, ± 0.004% FS
-12
± 0.05% rdg, ± 0.004% FS
-2
± 0.05% rdg, ± 0.004% FS
-6
± 0.05% rdg, ± 0.004% FS
-1
± 0.05% rdg, ± 0.004% FS
0
± 0.05% rdg, ± 0.004% FS
0
± 0.05% rdg, ± 0.004% FS
6
± 0.05% rdg, ± 0.004% FS
1
± 0.05% rdg, ± 0.004% FS
12
± 0.05% rdg, ± 0.004% FS
2
± 0.05% rdg, ± 0.004% FS
18
± 0.05% rdg, ± 0.004% FS
3
± 0.05% rdg, ± 0.004% FS
24
± 0.05% rdg, ± 0.004% FS
4
± 0.05% rdg, ± 0.004% FS
30
± 0.05% rdg, ± 0.004% FS
5
± 0.05% rdg, ± 0.004% FS
SELECT OPTION
CALIBRATION
PRESSURE INT
VOLTAGE IN
CURRENT OUT
TEMPERATURE
CURRENT IN
SET VALUE
Store zero point?
Apply nominal zero value
Press ENTER key when
applied value is stable
CALIBRATING VOLTAGE
RANGE +/- 30V
CALIBRATING VOLTAGE
RANGE +/- 30V
Enter Applied Value:
CHANGE
VALUE
RUN
CALIBRATING VOLTAGE
RANGE +/- 30V
YES
NO
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......