Copyright 2007 Baker Hughes Company.
14 | DPI 610/615 IS Instruction Manual–English
Chapter
4.
Taking Measurements
•
If applicable, turn on Cal Mode and set-up test parameters as detailed in Section
3.4.
4.3 Pressure Converter (Pressure to Pressure) Task
Select the P-P task from the task menu and connect the Unit Under Test (UUT) to the calibrator
as shown below. Testing a converter requires one pressure to be applied to the unit under test
(UUT) and another (converter output) to be measured. The additional measurement is provided
by the external transducer option.
•
Connect the UUT to the instrument as shown below. Plug the external transducer into the
instrument as shown below:
•
Press the
TASK
key and select the P-P task. Providing the external transducer has been
calibrated and its parameters stored in the instrument, the display will show
External
pressure in the input window and calibrator
Output
pressure in the output window. If an error
message “NO SENSOR OR CAL INVALID” is displayed, this indicates that the external
transducer has not been entered and/or calibrated with the instrument. Refer to Section
for details of adding an external transducer.
Note:
If an external transducer change is made, switch the calibrator off and then on to load
new transducer data.
•
If required, select the output units as described in Section
•
If applicable, turn on Cal Mode and set-up test parameters as detailed in Section
3.4.
A
External pressure source (indicator instruments only)
B
Pressure regulator
C
External pressure source
D
Pressure to pressure device
E
External supply
3
A
B
C
D
E
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......