Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 23
Process Functions
Note:
Display shows the last entered Tare Value.
5.4.2 Min/Max Process Function
To set-up an input display to show min/max and present input reading, enable MIN/MAX from the
process menu and press F1 (SETTINGS) to provide
RESET
function. The display now shows the
max/min values as follows:
Reset Max/Min display at any time by pressing the F1 key.
To quit max/min, press
INPUT
, select
MAX/MIN
from process menu and switch the function off.
5.4.3 Filter Process Function
To apply the low pass filter to a selected input, enable
FILTER
from the process menu and press
F1 (SETTINGS) to provide access to the filter parameters. Two settings are required, Time to
Settle and Band.
To examine the current filter settings and exit without change, press the
EXIT
key.
The set-up procedure is as follows:
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
PRESSURE INT
MAKE READING
ZERO
TARE OFF
A VALUE
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
PRESSURE INT
ENTER VALUE
VALUE: 10.000
NEW VALUE: _
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
PRESSURE INT
RESET
MAX/MIN
MAXIMUM
PRESENT
MINIMUM
SET FILTER
SELECT VALUE
SELECT VALUE
CHANGE
VALUE
CHANGE
VALUE
TIME TO SETTLE
TIME =
2
Seconds
TIME TO SETTLE
TIME = 2 Seconds
SET FILTER
BAND = MAX AMPLITUDE
OF NOISE TO BE
FILTERED, CHANGES
GREATER THAN BAND
BYPASS THE FILTER
BAND = 1.5 % FS
BAND = MAX AMPLITUDE
OF NOISE TO BE
FILTERED, CHANGES
GREATER THAN BAND
BYPASS THE FILTER
BAND =
1.5
% FS
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......