26 | DPI 610/615 IS Instruction Manual–English
Chapter
5.
Advanced Task
•
Press
STOP
(F1) when running to stop at any point. Press
RUN
(F1) to resume.
5.6.2 mA Ramp
Press the
OUTPUT
key and select
mA
Ramp
as shown previously in mA Step.
•
Define ramp required by entering
START
and
END
current values as shown below:
•
Connect an external power source as shown in Section
5.6.
•
Press
START
(F1) to run the program. A status display
CHECK LOOP
indicates a fault in
the external loop i.e. supply fault or open circuit.
Note:
The ramp cycle (min to max or max to min), is approximately 60 seconds.
•
Press
STOP
(F1) when running to stop at any point. Press
CONTINUE
(F1) to resume from
point of pause or
RESET
(F2) to return to start point.
5.6.3 mA Value
Press the
OUTPUT
key and select
mA
Value
from the output menu. The procedure is shown
below:
•
Connect an external power source as shown in Section
5.6.
TASK : ADVANCED
SNAPSHOT MODE
PRESSURE INT
TASK : ADVANCED
SNAPSHOT MODE
PRESSURE INT
START
OUTPUT
END
CURRENT
START
OUTPUT
END
CURRENT
START
START
STOPPED
STOPPED
ENTER VALUE
ENTER VALUE
TASK : ADVANCED
SNAPSHOT MODE
PRESSURE INT
TASK : ADVANCED
SNAPSHOT MODE
PRESSURE INT
START
OUTPUT
END
CURRENT
START
OUTPUT
END
CURRENT
START
STOP
STOPPED
RAMPING
ENTER VALUE
ENTER VALUE
TASK : ADVANCED
SNAPSHOT MODE
PRESSURE INT
TASK : ADVANCED
SNAPSHOT MODE
PRESSURE INT
START
OUTPUT
END
CURRENT
START
OUTPUT
END
CURRENT
STOP
CONTINUE
RESET
STOPPED
PAUSED
ENTER VALUE
ENTER VALUE
PRESSURE INT
PRESSURE EXT
mA STEP
mA RAMP
mA VALUE
SELECT OUTPUT
SELECT OUTPUT
SET LEVEL
TASK : ADVANCED
SNAPSHOT MODE
CURRENT
PRESSURE INT
OK
OUTPUT
NEW VALUE
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......