6. Memory Operations
6.1 Saving Display or Data Log
In Store Mode three memory operations can be set-up:
None
,
Snapshot
and
Datalog
. Refer to
7.2 for details.
6.2 Store Operations (Screen Snapshots)
To store any display (menu displays excepted), press the
STORE
key. This saves the current
display to the next available location. Supporting text (10 characters) may be appended. Twenty
memory locations are available on a cyclic buffer. When all 20 have been used, store operations
overwrite existing locations, starting at Location 1.
6.3 Recalling Stored Data (Screen Snapshots)
To recall a previously stored display, press the
RECALL
key. This recalls the last display saved.
Press the
left
or
right
cursor keys to recall the previous or next locations respectively. To
exit
RECALL
, press the
EXIT
key.
6.4 Data Log Operations
Data Log is a special application of store mode which enables the calibrator to either
automatically log displays at preset time intervals or to manually log a display on operation of the
STORE
key. Logged data is written to a user specified file.
To set-up a Data Log file, proceed as follows.
•
Select a task, other than
BASIC
. If using
ADVANCED
, set-up the required output
parameters.
•
Use
SETUP
to select Data Log from the Store Mode Menu (See Section
7.2).
TASK : ADVANCED
SNAPSHOT MODE
PRESSURE INT
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
ENTER TEXT
PRESSURE INT
TEXT :
SNAPSHOT 19
TASK : ADVANCED
SNAPSHOT 2 TEST19
PRESSURE INT
VOLTAGE
SELECT LOCATION
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......