Pressure to Display (P-DISPLAY) Task
4.6 Pressure to Display (P-DISPLAY) Task
P-Display is a special application of Data Log. To use this mode, select Data Log from the Store
Mode menu as detailed in Section
7.2. Connect the UUT to the instrument as shown below and,
if required, turn on and set-up Cal Mode (see Section
3.4).
•
Press
TASK
and select
P-DISPLAY
. If required, use
OUTPUT
key to change pressure units.
•
Set-up a data log file as detailed in Section
6.4.
Note: TRIGGER
field, automatically set to
KEYPRESS
, cannot be changed.
•
Apply a series of test pressures to the device under test. Enter displayed leading at each
pressure and log each point:
•
After logging final test point, terminate as follows:
A
External pressure source
B
Pressure regulator
C
Dial gauge under test
PRESSURE INT
DISPLAY
CHANGE
VALUE
TASK : P-DISPLAY
SNAPSHOT MODE