Copyright 2007 Baker Hughes Company.
50 | DPI 610/615 IS Instruction Manual–English
Chapter
8.
Calibration
8. Connect an external voltage source shown in Figure
49, press the
OUTPUT
key, select
mA
Value
:
9. Set the loop output current to 5 mA as shown below and check that the output current,
measured by the ammeter standard is within the limits given in Table
50.
10. Repeat (9) for all output values given in Table
50.
11. Disconnect the calibration/test equipment.
8.14 Ambient Temperature Channel
Use the following procedure for calibrating the ambient temperature measurement channel:
Table
8
-
5: Electrical Current Output Calibration Tolerances
Current Output Calibration, 24 mA Range
Verification Tolerances
Applied Current
(mA)
Permissible Deviation
0
± 0.05% rdg ± 0.01% FS
5
± 0.05% rdg ± 0.01% FS
10
± 0.05% rdg ± 0.01% FS
15
± 0.05% rdg ± 0.01% FS
20
± 0.05% rdg ± 0.01% FS
24
± 0.05% rdg ± 0.01% FS
20
± 0.05% rdg ± 0.01% FS
15
± 0.05% rdg ± 0.01% FS
10
± 0.05% rdg ± 0.01% FS
5
± 0.05% rdg ± 0.01% FS
0
± 0.05% rdg ± 0.01% FS
PRESSURE INT
mA STEP
mA RAMP
mA VALUE
SELECT OUTPUT
UNITS
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
PRESSURE INT
PRESSURE EXT
SELECT OUTPUT
OUTPUT
NEW VALUE
CURRENT
OK
VOLTAGE
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
TASK : ADVANCED
SNAPSHOT MODE
OUTPUT
NEW VALUE
CURRENT
OK
OUTPUT
NEW VALUE
CURRENT
OK
OUTPUT
NEW VALUE
CURRENT
OK
SELECT OUTPUT
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......