6.
Memory Operations
6.4.5 Delete Datalog and Procedure Files
To delete a Data Log file, or a procedure file (DPI 615 instruments only), proceed as follows.
Alternatively, to delete all logged files simultaneously, select
ERASE ALL FILES
(F2) at the erase
screen.
6.4.6 Downloading Procedure Files
Complete test procedures may be downloaded from a PC to the instrument (DPI 615 instruments
only) via the RS232 port. A procedure consists of a number of Druck Universal Command
Interface (DUCI) commands that are usually assembled by a linking management software
application (e.g.)
4Sight™ 2
.
Before downloading a procedure, the instrument must be in the REMOTE mode. To place the
instrument into REMOTE mode, proceed as follows.
•
Connect the instrument’s RS232 port to a free COM port on a PC.
•
Make sure that the COMMS parameters of the PC match those of the instrument (refer to
7.4.4).
•
Make sure that the instrument is not already running a procedure. If it is, quit the procedure.
•
Download the procedure. Procedures are stored in the Data Log directory.
The following sequence shows a typical download sequence that starts with the instrument in
LOCAL mode.
After the procedure file has been downloaded, the instrument is usually returned to the LOCAL
mode by the last command in the procedure file. If the instrument remains in REMOTE mode,
switch it OFF and ON to reset it.
WARNING
The RS232 interface must only be used in a safe area.
A
Local mode
B
Remote mode (Download)
C
Local (auto)
MEMORY USED
FILE 1 of 9
DATE: 04/11/99 TIME 10:18
ERASE
SELECT FILE
MEMORY USED
FILE 1 of 9
DATE: 04/11/99 TIME 10:18
ERASE
ALL FILES
ERASE
CHOSEN FILE
MEMORY USED
FILE 1 of 9
DATE: 04/11/99 TIME 10:18
ERASE CHOSEN FILE?
YES
NO
TASK : ADVANCED
SNAPSHOT MODE
TASK : ADVANCED
SNAPSHOT MODE
TASK : ADVANCED
SNAPSHOT MODE
VOLTAGE
PRESSURE INT
SETTINGS
VOLTAGE
PRESSURE INT
VOLTAGE
PRESSURE INT
SETTINGS
DOWNLOAD
SUCCESSFUL
REMOTE
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......