Copyright 2007 Baker Hughes Company.
52 | DPI 610/615 IS Instruction Manual–English
Chapter
8.
Calibration
3. Switch the instrument on, enter
CALIBRATION
mode and select
EXT
SENSORS
from the
CALIBRATION
menu:
4. Select the transducer to be calibrated from the transducer menu using the cursor keys and
press ENTER:
If the sensor to be calibrated is not in the directory, or no sensors are listed in the directory,
press the
F1
key (ADD NEW SENSOR). This will place the sensor in the directory, allowing
it to then be selected.
5. Connect the pressure standard to the inlet of the external transducer, apply the zero point
pressure and store the zero point:
6. Apply the full-scale pressure to the external transducer and store the full-scale (FS) point
7. Release the applied pressure and disconnect the pressure reference. Press the
ENTER
key
to accept the calibration. Press the
EXIT
key once to return to the calibration menu or four
times to quit the
CALIBRATION
and
SETUP
modes.
8. Check the calibration of the external transducer by applying the test pressures as detailed in
INT RANGES
CHANGE PIN
EXT SENSORS
CALIBRATION
SELECT CAL OPTION
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
Apply nominal zero value
Press ENTER key when
applied value is stable
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
Enter Applied Value:
CHANGE
VALUE
RUN
SET VALUE
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
YES
NO
Store zero point?
+
CHANGE VALUE
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
Apply nominal FS value
Press ENTER key when
applied value is stable
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
Enter Applied Value:
CHANGE
VALUE
RUN
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
Enter Applied Value:
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
Store FS point?
YES
NO
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
RESULTS
Calibration Complete
Press ENTER to continue
CHANGE
VALUE
Applied
Measured
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......