Copyright 2007 Baker Hughes Company.
64 | DPI 610/615 IS Instruction Manual–English
Appendix
B.
“Quick Fit” Pressure Connector
b. BSP (parallel type, 100 bar (1500 psi) or more): We recommend a bonded seal at the
bottom.
c. BSP (parallel type, 100 bar (1500 psi) or less): a bonded seal at the top is permitted).
3. Attach the adaptor to the device. If necessary, use one of the alternative adaptor sets, part
number IO620-BSP, IO620-NPT or IO620-MET, and tighten to the applicable torque.
4. Re-attach the adaptor to the “Quick Fit” pressure connector and hand tighten it.
Figure
B
-
3: Remove/Install “Quick Fit” Pressure Connector
Procedure to remove the “Quick Fit” pressure connector:
1. Release the pressure.
2. Remove the adaptor from the “Quick Fit” pressure connector. The adaptor is hand tightened
and does not require tools to remove.
3. Remove the adaptor from the device under test; use appropriate tools as necessary.
4. Re-attach the adaptor to the “Quick Fit” pressure connector and hand tighten it.
a. NTP
b. (G)/BSPP
P
EDU
c. (G)/BSPP
3EDU
3UHVVXUH3RUW