English–DPI 610/615 IS Instruction Manual | 49
Current Output Range (24 mA)
Note:
This procedure requires an external supply.
Figure
8
-
4: 24 mA Current Output Range Calibration Circuit
4. Measure the 10% full-scale output current and enter the value measured on the external
ammeter:
5. Measure the 90% full-scale output current and enter the value measured on the external
ammeter
6. Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to quit the
CALIBRATION
and
SETUP
modes.
7. Verify the instrument’s output current calibration by setting up the loop test currents as
shown in Table
50 and checking the set values against the ammeter standard.
55mA
Vin
mA in
mA Sink
max 30V
max
mA
-
+
V
CALIBRATING CURRENT
RANGE 0-24 mA
Measure 10% FS output
Press ENTER key when
applied value is stable
CALIBRATING CURRENT
RANGE 0-24 mA
Enter Measured Current:
CALIBRATING CURRENT
RANGE 0-24 mA
Enter Measured Current:
CHANGE
VALUE
RUN
CHANGE
VALUE
CHANGE VALUE
CALIBRATING CURRENT
RANGE 0-24 mA
CALIBRATING CURRENT
RANGE 0-24 mA
Measure 90% FS - output
Press ENTER key when
applied value is stable
CALIBRATING CURRENT
RANGE 0-24 mA
Enter Measured Current:
CHANGE
VALUE
RUN
CALIBRATING CURRENT
RANGE 0-24 mA
Enter Measured Current:
CHANGE VALUE
CHANGE
VALUE
CALIBRATING CURRENT
RANGE 0-24 mA
Store Cal point?
YES
NO
Store Cal point?
YES
NO
CALIBRATING CURRENT
RANGE 0-24 mA
RESULTS
Calibration Complete
Press ENTER to continue
Applied
Measured
Summary of Contents for Druck DPI 610 IS
Page 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Page 2: ......
Page 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Page 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Page 75: ......