286
Maintaining Your GC
14
Maintaining the NPD
To Bakeout the NPD
1
Bakeout the NPD with the column installed or
uninstalled. If uninstalled, gather the following (see
“Consumables and Parts for the NPD”
•
Capillary adapter (adaptable NPD only)
•
Column nut
•
No- hole ferrule
2
Load the
and wait for the GC to
become ready.
3
Set the bead voltage to 0, then turn it Off.
4
If the column is uninstalled, plug the detector connection
with the capillary adapter, column nut, and no- hole
ferrule.
Maintain inert carrier gas flow through the column, or
remove the column from the GC.
5
Set normal operating flows.
6
Set the detector temperature at 25 °C above the typical
method setpoint temperature.
7
Set the oven temperature to 250 °C or 25 °C above the
normal maximum operating temperature. Do not exceed
the column’s temperature limit.
8
Hold at temperature for 15 to 30 minutes.
9
If the column is not installed in the NPD, install it. (See
“To Install a Capillary Column in the NPD”
10
Restore the analytical method and allow the NPD to
equilibrate at operating temperatures and flows for 10 to
30 minutes.
11
Check the NPD leakage current. See
WA R N I N G
If using hydrogen as a carrier gas, turn off the hydrogen supply
and cap the end of the column to prevent an oven explosion.
WA R N I N G
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...