![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 201](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996201.webp)
Maintaining the FID
11
Maintaining Your GC
201
Selecting an FID Jet
Open the oven door and locate the column connection fitting
at the base of the detector. It will look like either a capillary
optimized fitting or an adaptable fitting.
•
If you have an application that tends to clog the jet, select
a jet with a wider tip id.
•
When using packed columns in high column- bleed
applications, the jet tends to clog with silicon dioxide.
•
In simulated distillation applications, the high- boiling
hydrocarbons tend to clog the jet.
For capillary optimized fittings, select a jet from
.
For adaptable fittings, select a jet from
Capillary optimized fitting
Adaptable fitting
Detector fitting
Adapter
Table 32
Jets for capillary optimized fittings
Figure 1 ID
Jet type
Part number
Jet tip id
Length
1
Capillary
G1531-80560
0.29 mm (0.011 inch)
43 mm
2
High-temperature
(use with simulated
distillation)
G1531-80620
0.47 mm (0.018 inch)
48 mm
Figure 3
Capillary optimized jets
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...